CivaTech Oncology Inc., CT40 CivaThin

OptiSeed_1032P.png

Source Description:

The dimensions for the CT40 CivaThin 1, 2  are taken from the studies by Rivard et al 1 and Reed et al 2. The CivaThin family sources are available in lengths of 10, 20, 30, 40, and 50 mm, and referred to as CT10 to CT50, respectively. The CivaThin sources are identical to the CivaString sources, except their outer diameter is smaller. Encapsulation for the CT40 CivaThin source is made of organic polymer material with a mass density of 1.15 g/cm3  and weight fractions of 8.5%, 64.8%, 5.0%, and 21.7%, of H, C, N, and O, respectively.  The outer diameter of the source is 0.60 mm. The ends are sealed as a solid cylinder (diameter 0.60 mm). The CT40 source consist of four cylindrical Au markers with a diameter of 0.250 mm and a length of 0.80 mm. They are off center at ± 0.5 cm and ± 1.51 mm along the length of the source with a  0.0550 mm distance offset between the cylindrical axis of the Au markers and the cylindrical axis of the organic polymer Overall the CT40 source includes sixteen Pd regions. Two 103Pd regions are located on either side of each Au marker (overall sixteen regions) centered at longitudinal distances of ± 0.111 mm, ± 0.333 mm, ± 0.667 mm, ± 0.889 mm, ± 1.121 mm, ± 1.343 mm, ± 1.670 mm, and ± 1.889 mm offset from the source center. Each region is simulated similar to an isoceles trapezoidal prism with a long and short base with a length of 2.10 mm and 1.940 mm, and a height and a width of 0.250 mm and 0.280 mm, respectively. Each region has a rectangular cross section, and there is a 0.0350 mm distance between the longitudinal axis of each region and central axis of the organic polymer. Each Pd region is modeled with a 5% weight fraction of natural Pd (~8 μg of Pd) and 95% of organic polymer (H=0.08075, C=0.6156, N=0.0475, O=0.20615,  Pd=0.05). The Au markers and each well are fixed in organic polymer encapsulation without any air space. The overall source length is 40.90 mm and the active length is 40.0 mm. The mean photon energy calculated on the surface of the source is 20.58 keV with mean statistical uncertainties < 0.01% .


Dose-Rate Constant - Λ :

Dose-rate constants, Λ , are calculated by dividing the dose to water per history in a (0.1 mm)3 voxel centered at the reference position,         (1 cm,Π/2), in the 30x30x30 cm3  water phantom, by the air-kerma strength per history (scored in vacuo either in a detector 2.363x2.363x0.001 cm3 which subtends the same solid angle of the NIST WAFAC detector or in a small point detector, 0.1x0.1x0.001 cm3). Paxton et al 3 show that, for geometrical reasons, the NIST WAFAC cannot directly measure the air-kerma strength for long sources (> 1 cm) because some photons would create charge outside the charge collecting region. Nonetheless, assuming WAFAC or VAFAC measurements deliver the air-kerma averaged over their front surface, kr2 corrections are necessary to convert the average kerma on the WAFAC/VAFAC detector’s front surface to the kerma on the axis, as required for determining the air-kerma strength. We have determined kr2 corrections by calculating the average kerma per history in a thin (0.001 cm) circular (radius = 1.333 cm)  detector/voxel located 10 cm from the source's axis for a series of point sources off-axis and comparing to the kerma per history in a small point detector (0.1x0.1x0.001 cm3) on the axis. The detector covered the same solid angle as the NIST WAFAC 4, 5 or UW VAFAC 3 detectors at 30 cm. The values of kr2 for line sources are then determined by integrating the corrections as a function of the point source distance off-axis. The values obtained are 1.0094,1.0092, 1.0090, 1.0086, and 1.0084 for a point source, 1 cm, 2 cm, 4 cm and 5 cm line sources, respectively. For a point source on axis these values are comparable with the analytic value of 1.0094 (square detector) or 1.0090 (circular detector) 6 .   MC uncertainties in the table are statistical uncertainties only (k=1).

Author Method Λ (cGy h-1 U-1) Abs. Uncertainty
Safigholi et al 7 WAFAC 0.33124 0.00004
Safigholi et al 7 Point 0.33105 0.00015
Rivard 1 WAFAC (MCNP) 0.328 0.004

 Radial dose function - g(r):

The radial dose function, g(r), is calculated using both line and point source geometry functions and tabulated at 36 different radial distances ranging from 0.19 cm to 10 cm. Fit parameters for a modified polynomial expression are also provided 7. The mean residual deviations from the actual data for the best fit were < 0.44%.  The supplementary data of Ref. 1 are used in the  g(r), and F(r, Θ) comparisons.

Click image for higher res version

radial dose function

Fitting coefficients for g L (r) = (a0 r-2 + a1 r-1 + a2 + a3r + a4r2 + a5 r3) e-a6r
Fit range Coefficients
r min (cm) r max (cm)
0.05 10.00 a0 / cm2 (-1.56+/-0.05)E-03
    a1 / cm (3.76+/-0.11)E-02
    a2 (1.636+/-0.005)E+00
    a3 / cm-1 (1.66+/-0.14)E-01
    a4 / cm-2 (-4.19+/-0.10)E-02
    a5 / cm-3 (2.53+/-0.05)E-03
    a6 / cm-1 (5.79+/-0.06)E-01

Anisotropy function - F(r,θ):

Anisotropy functions are calculated using the line source approximation and tabulated at radii of 0.1, 0.15, 0.25, 0.5, 0.75, 1, 2, 3, 4, 5, 7.5 and 10 cm and 32 unique polar angles with a minimum resolution of 5° . The anisotropy factor, φan (r), was calculated by integrating the solid angle weighted dose rate over 0° ≤ ϑ ≤ 90° .

Click images for higher res versions
F(0.1,θ)
Anisotropy function
F(1.00,θ)
Anisotropy function

F(5.00,θ)
Anisotropy function

 F(7.50,θ)
Anisotropy function

 

Tabulated data: 

Tabulated data are available in .xlsx format: Excel


   References:

1. M. J. Rivard et al , 103Pd strings: Monte Carlo assessment of a new approach to brachytherapy source design, Med. Phys., 41 , 011716-11, 2014.
2. J. L. Reed et al , Experimental  and Monte Carlo dosimetric characterization of a 1 cm 103Pd brachytherapy source, Brachytherapy, 13 , 657-667, 2014. 
3. A.B. Paxton, W.S. Culberson, L.A. DeWerd, and J.A. Micka, Primary calibration of coiled brachytherapy sources, Med. Phys. 35, 32-38,2 008 
4. R. Loevinger, Wide-angle free-air chamber for calibration of low-energy brachytherapy sources, Med. Phys., 20 , 907, 1993. 
5. S. M Seltzer et al , New National Air-Kerma-Strength Standards for 125I and 103Pd Brachytherapy Seeds, J. Res. Natl. Inst. Stand. Technol.,108,337-358,2003.                                                                                                                                                                              6. D. W. O. Rogers, Inverse square corrections for FACs and WAFACs, Appl. Radiat. Isot. 153, 108638 (8pp), 2019.                                        7. H. Safigholi, M. J. P. Chamberland, R. E. P. Taylor, C. H. Allen, M. P. Martinov, D. W. O. Rogers, and R. M. Thomson, Update of the CLRP TG-43 parameter database for  brachytherapy,  to be published (Current calculation).
8. M. J. Rivard et al ,  Supplement 2 for the 2004 update of the AAPM Task Group No. 43 Report: Joint recommendations by the AAPM and GEC-ESTRO, Med. Phys., 44 , e297-e338, 2017.                                                                                                                                            
9. R. E. P. Taylor, D. W. O. Rogers, More accurate fitting of 125I and 103Pd radial dose functions, Med. Phys., 35 , 4242-4250, 2008.


Carleton Laboratory for Radiotherapy Physics 

 CLRP TG-43 Parameter Database V2
 May 5, 2020

Search Carleton