CivaTech Oncology Inc., CS20 CivaString

OptiSeed_1032P.png

Source Description:

The dimensions for the CS20 CivaString 1, 2 are taken from the studies by Rivard 1 and Reed et al 2. The CivaString family of sources are available in lengths of 10, 20, 30, 40, and 50 mm, and referred to as CS10 to CS50, respectively. The CivaString sources are identical to the CivaThin sources, except their outer diameter is a little larger. Encapsulation for the CS20 CivaString source is made of organic polymer material with a mass density of 1.15 g/cm3  and weight fractions of 8.5%, 64.8%, 5.0%, and 21.7%, of H, C, N, and O, respectively. The outer diameter of the source is 0.850 mm. The ends are sealed as a solid cylinder (diameter 0.850 mm). The CS20 source consists of two cylindrical Au markers with a diameter of 0.250 mm and a length of 0.80 mm located off centered at ± 0.5 cm along the length of the source with a  0.0550 mm distance offset between the cylindrical axis of the Au markers and the cylindrical axis of the organic polymer. Two 103Pd regions are located on either side of the Au markers (overall eight regions) centered at longitudinal distances of ± 0.111 mm, ± 0.333 mm, ± 0.667 mm,  and ± 0.889 mm offset from the source center. Each region is simulated similar to an isoceles trapezoidal prism with a long and short base with of lengths 2.10 mm and 1.940 mm, and a height and a width of 0.250 mm and 0.280 mm, respectively. Each region has a rectangular cross section, and there is a 0.0350 mm distance between the longitudinal axis of each region and the central axis of the organic polymer. Each Pd region is modelled with a 5% weight fraction of natural Pd (~8 μg of Pd) and 95% of organic polymer (H=0.08075, C=0.6156, N=0.0475, O=0.20615,  Pd=0.05). The Au markers and each region are fixed in the organic polymer encapsulation without any air space. The overall source length is 20.90 mm and the active length is 20.0 mm. The mean photon energy calculated on the surface of the source is 20.58 keV with statistical uncertainties < 0.01% .


Dose-Rate Constant - Λ :

Dose-rate constants, Λ , are calculated by dividing the dose to water per history in a (0.1 mm)3 voxel centered at the reference position,         (1 cm,Π/2), in the 30x30x30 cm3  water phantom, by the air-kerma strength per history (scored in vacuo either in a detector 2.363x2.363x0.001 cm3 which subtends the same solid angle of the NIST WAFAC detector or in a small point detector, 0.1x0.1x0.001 cm3). Paxton et al 3 show that, for geometrical reasons, the NIST WAFAC cannot directly measure the air-kerma strength for long sources (> 1 cm) because some photons would create charge outside the charge collecting region. Nonetheless, assuming WAFAC or VAFAC measurements deliver the air-kerma averaged over their front surface, kr2 corrections are necessary to convert the average kerma on the WAFAC/VAFAC detector’s front surface to the kerma on the axis, as required for determining the air-kerma strength. We have determined kr2 corrections by calculating the average kerma per history in a thin (0.001 cm) circular (radius = 1.333 cm) detector/voxel located at 10 cm from the axis for a series of point sources off-axis and comparing to the kerma per history in a small point voxel (0.1x0.1x0.001 cm3) on the axis. The detector covered the same solid angle as the NIST WAFAC 4, 5 or UW VAFAC 3 detectors at 30 cm. The values of kr2 for line sources is then determined by integrating the corrections as a function of the point source distance off-axis. The values obtained are 1.0094,1.0092, 1.0090, 1.0086, and 1.0084 for a point source, 1 cm, 2 cm, 4 cm, 5 cm line sources, respectively. For a point source on axis these valuse are comaprable with the analytic value of 1.0094 (square detector) or 1.0090 (circular detector) 6. MC uncertainties in the table are statistical uncertainties only (k=1).

Author Method Λ (cGy h-1 U-1) Abs. Uncertainty
Safigholi et al 7 WAFAC 0.51178 0.00005
Safigholi et al 7 Point 0.51163 0.00011
Rivard 1 WAFAC (MCNP) 0.510 0.007

 Radial dose function - g(r):

The radial dose function, g(r), is calculated using both line and point source geometry functions and tabulated at 36 different radial distances ranging from 0.19 cm to 10 cm. Fit parameters for a modified polynomial expression are also provided  8. The mean residual deviations from the actual data for the best fit were < 0.35%.  The supplementary data of Ref. 1 are used in the  g(r), and F(r, Θ) comparisons.

Click image for higher res version

radial dose function

Fitting coefficients for g L (r) = (a0 r-2 + a1 r-1 + a2 + a3r + a4r2 + a5 r3) e-a6r
Fit range Coefficients
r min (cm) r max (cm)
0.05 10.00 a0 / cm2 (-1.014+/-0.034)E-03
    a1 / cm (2.26+/-0.08)E-02
    a2 (1.589+/-0.004)E+00
    a3 / cm-1 (3.28+/-0.19)E-01
    a4 / cm-2 (-5.90+/-0.10)E-02
    a5 / cm-3 (3.70+/-0.08)E-03
    a6 / cm-1 (6.34+/-0.08)E-01

Anisotropy function - F(r,θ):

Anisotropy functions are calculated using the line source approximation and tabulated at radii of 0.1, 0.15, 0.25, 0.5, 0.75, 1, 2, 3, 4, 5, 7.5 and 10 cm and 32 unique polar angles with a minimum resolution of 5° . The anisotropy factor, φan (r), was calculated by integrating the solid angle weighted dose rate over 0° ≤ ϑ ≤ 90° .

Click images for higher res versions
F(0.10,θ)
Anisotropy function
F(0.50,θ)
Anisotropy function

F(1.00,θ)
Anisotropy function

 F(3.00,θ)
Anisotropy function

F(5.00,θ)
Anisotropy function

 F(7.50,θ)
Anisotropy function


Tabulated data: 

Tabulated data are available in .xlsx format: Excel


  References:

1. M. J. Rivard et al , 103Pd strings: Monte Carlo assessment of a new approach to brachytherapy source design, Med. Phys., 41 , 011716-11, 2014.
2. J. L. Reed et al , Experimental  and Monte Carlo dosimetric characterization of a 1 cm 103Pd brachytherapy source, Brachytherapy, 13 , 657-667, 2014. 
3. A.B. Paxton, W.S. Culberson, L.A. DeWerd, and J.A. Micka, Primary calibration of coiled brachytherapy sources, Med. Phys. 35, 32-38,2 008 
4. R. Loevinger, Wide-angle free-air chamber for calibration of low-energy brachytherapy sources, Med. Phys., 20 , 907, 1993. 
5. S. M Seltzer et al , New National Air-Kerma-Strength Standards for 125I and 103Pd Brachytherapy Seeds, J. Res. Natl. Inst. Stand. Technol.,108,337-358,2003.                                                                                                                                                                              6. D. W. O. Rogers, Inverse square corrections for FACs and WAFACs, Appl. Radiat. Isot. 153, 108638 (8pp), 2019.                                        7. H. Safigholi, M. J. P. Chamberland, R. E. P. Taylor, C. H. Allen, M. P. Martinov, D. W. O. Rogers, and R. M. Thomson, Update of the CLRP TG-43 parameter database for  brachytherapy,  to be published (Current calculation).
8. R. E. P. Taylor, D. W. O. Rogers, More accurate fitting of 125I and 103Pd radial dose functions, Med. Phys., 35 , 4242-4250, 2008.                                                                                                                                          