Draximage, BrachySeed, LS-1

BrachySeed_LS-1.png

Source Description:

Seed dimensions for the LS-1 seed   are the same as those used in the study by Williamson 1. The DRAXIMAGE BrachySeed LS-1 consists of two 0.5mm diameter aluminum silicate spheres containing a uniform distribution of 125I. Density of the aluminum silicate is 2.81 g/cm 3 and composition by weight is 40.7% oxygen, 21.4% silicon, 16.6% aluminum, 11.3% sodium and 10.0% silver (the amount of 125I within the spheres is assumed to be negligible). The two source spheres are separated by a 90% Pt / 10% Ir cylindrical rod that is 3.0 mm long and has a diameter of 0.380 mm. At the middle of the seed is a 1.20 mm long Ti annulus with inner and outer diameters of 0.390 mm and 0.698 mm, respectively. The wall thickness of the cylindrical portion of the source is 0.10 mm. The hemi-spherical ends are 0.065 mm thick on the longitudinal axis and 0.050 mm thick where they meet the cylindrical walls. End welds are modelled using a 0.40 mm radius Ti hemisphere overlapped with a 0.350 mm radius air sphere with its center shifted by 0.015 mm relative to the Ti sphere. The overall source length is 4.40 mm and the active length is 4.10 mm. The two spheres are free to move approximately 0.050 mm along the seed axis and 0.050 mm radially. The mean photon energy calculated on the surface of the source is 27.83 keV with statistical uncertainties < 0.011%

Dose-Rate Constant - Λ :

Dose-rate constants, Λ , are calculated by dividing the dose to water per history in a (0.1 mm)voxel centered on the reference position,        (1 cm,Π/2), in the 30x30x30 cmwater phantom, by the air-kerma strength per history (scored in vacuo). As described in ref. 2 , dose-rate constants are provided for air-kerma strength calculated using voxels of 2.66x2.66x0.05 cm3 (WAFAC) and 0.1x0.1x0.05 cm3 (point) located 10 cm from the source. The larger voxel size averages the air-kerma per history over a region covering roughly the same solid angle subtended by the primary collimator of the WAFAC 3, 4  at NIST used for calibrating low-energy brachytherapy sources and is likely the most clinically relevant value. The small voxel serves to estimate the air kerma per history at a point on the transverse axis and includes a small 1/r2 correction (0.5%) 2. egs_brachy and BrachyDose MC uncertainties are only statistical uncertainties (k=1).

Author Method Λ (cGy h-1 U-1) Abs. Uncertainty
Safigholi et al 5 WAFAC 0.9250 0.0001
Safigholi et al 5 point 0.9230 0.0018
Taylor, Rogers 6 WAFAC 0.922 0.002
Taylor, Rogers 6 point 0.922 0.005
Williamson 1 WAFAC (PTRAN) 0.935 0.017
Wang, Sloboda 7 Point (EGS4) 0.932 0.003
Nath, Yue 8 TLD 1.02 0.07
Rodrigues, Rogers 9 TLD (Revised Nath) 0.943 0.048
Rodrigues, Rogers 9 WAFAC (BrachyDose) 0.922 0.015
Chan, Prestwich 10 GaF Chromic film 0.98 0.06
Chan, Prestwich 10 extrap (CYLTRAN) 0.90 0.03
Rivard et al 11 Consensus value 0.972  

Radial dose function - g(r):

The radial dose function, g(r), is calculated using both line and point source geometry functions and tabulated at 36 different radial distances ranging from 0.05 cm to 10 cm. Fit parameters for a modified polynomial expression are also provided 12 . The mean residual deviations from the actual data for the best fit were < 0.73%.  (Make marker of williamson larger to be visible)

Click image for higher res version

radial dose function

Fitting coefficients for g L (r) = (a0 r-2 + a1 r-1 + a2 + a3r + a4r2 + a5 r3) e-a6r
Fit range Coefficients
r min (cm) r max (cm)
0.10 10.0 a0 / cm2 (9.11+/-0.21)E-03
    a1 / cm (-0.2+/-1.1)E+00
    a2 (0.1+/-2.8)E+01
    a3 / cm-1 (0.0+/-1.8)E+02
    a4 / cm-2 (0+/-4)E+01
    a5 / cm-3 (0+/-5)E+00
    a6 / cm-1 (0.0+/-1.3)E+02
 

Anisotropy function - F(r,θ):

Anisotropy functions are calculated using the line source approximation and tabulated at radii of 0.1, 0.15, 0.25, 0.5, 0.75, 1, 2, 3, 4, 5, 7.5 and 10 cm and 32 unique polar angles with a minimum resolution of 5 o . The anisotropy factor, φ an (r), was calculated by integrating the solid angle weighted dose rate over 0 o ≤ ϑ ≤ 90 o .

Click images for higher res versions
F(0.25,θ)
Anisotropy function
F(0.50,θ)
Anisotropy function
F(1.00,θ)
Anisotropy function
F(5.00,θ)
Anisotropy function
 

Tabulated data:

Tabulated data are available in .xlsx format: Excel


References:

1. J. F. Williamson, Dosimetric characteristics of the DRAXIMAGE model LS-1 interstitial brachytherapy source design: A Monte Carlo investigation, Med. Phys., 29 , 509 - 521, 2002
2. R. E. P. Taylor et al , Benchmarking BrachyDose: voxel-based EGSnrc Monte Carlo calculations of TG-43 dosimetry parameters, Med. Phys., 34 , 445 - 457, 2007 
3. R. Loevinger, Wide-angle free-air chamber for calibration of low-energy brachytherapy sources, Med. Phys., 20 , 907, 1993 
4. S. M Seltzer et al , New National Air-Kerma-Strength Standards for 125I and 103Pd Brachytherapy Seeds, J. Res. Natl. Inst. Stand. Technol.,108,337-358,2003 
5. H. Safigholi, M. J. P. Chamberland, R. E. P. Taylor, C. H. Allen, M. P. Martinov, D. W. O. Rogers, and R. M. Thomson, Update of the CLRP TG-43 parameter database for LDR brachytherapy sources,  to be published (Current calculation).
6. R. E. P. Taylor, D. W. O. Rogers, An EGSnrc Monte Carlo-calculated database of TG-43 parameters, Med. Phys., 35 , 4228-4241, 2008 7. R. Wang, R. Sloboda, Monte Carlo dose parameters of the BrachySeed model LS-1 125I brachytherapy source, Appl. Radiat. Isotopes, 56 , 805-813, 2002 
8. R. Nath, N. Yue, Experimental determination of a newly designed encapsulated interstitial brachytherapy source of iodine-125-model LS-1 BrachySeed, Appl. Radiat. Isotopes, 55 , 813-821, 2001                                                                                                                     
9. M. Rodriguez , D. W. O. Rogers, Effect of improved TLD dosimetry on the determination of dose rate constants for 125I and 103Pd brachytherapyseeds,Med.Phys.41, 114301-15, 2014                                                                                                                                            10. G. Chan, W. V. Prestwich, Dosimetric properties of the new 125I BrachySeed model LS-1 source, Med. Phys., 29 , 190--200, 2002   
11. M. J. Rivard et al , Supplement to the 2004 update of the AAPM Task Group No. 43 Report, Med. Phys., 34 , 2187 - 2205, 2007
12. R. E. P. Taylor, D. W. O. Rogers, More accurate fitting of 125I and 103Pd radial dose functions, Med. Phys., 35 , 4242-4250, 2008