E&Z BEBIG GmbH, Co-60 HDR, Co0.A86

microSelectron_v1.png

Source Description:

Dimensions for the BEBIG Multisource 60Co HDR Co0.A86 source are taken from the study by Granero et al 1,2. The source consists of a 3.50 mm long cobalt core with a diameter of 0.5 mm enclosed in a stainless steel capsule (density 8.0 g/cm3, elemental composition: 63% Fr,  19% Cr, 12% Ni, 3% Mo, 2% Mn) of diameter 1.0 mm.  The tip of the encapsulation is assumed to be a hemisphere of radius 0.50 mm with its centre shifted 2.0 mm from the source centre. The core is housed in a hollow cylinder of air with an outer diameter of 0.7 mm. The air cylinder extends 0.06 mm from the core on the distal end of the source along the long axis. The air cylinder is capped with a conical section with a 118° opening angle. The encapsulation of the source has a total length of 5 mm. A stainless steel cable of length 5 mm and diameter 0.9 mm (density 5.75 g/cm3, elemental composition: 67% Fe, 17% Cr, 11% Ni, 2% Mo, 2% Mn) is attached to the source. The active length of this source is 3.5 mm. The mean photon energy calculated on the surface of the source is 1240.48 keV with statistical uncertainties < 0.001% .

Note:  All the results presented in the Table, Figures, xlsx, and zip files are with electron transport in phantom except where noted.


 Dose Rate Constant - Λ :

Dose-rate constants, Λ , are calculated by dividing the dose to water per history in a (0.1 mm)3 voxel centered on the reference position,   (1 cm, Π/2), in the 80x80x80 cm3 water phantom, by the air-kerma strength per history factor (scored in vacuo). Air kerma per history is always calculated using a tracklength estimator in a 10x10x0.05 cm3 air voxel located in vacuo on the transverse axis 100 cm away from the source and then corrected (kr2 = 1.00217) for the lateral and thickness dimensions of the scoring voxel to give the air kerma per history on the central axis at a point 100 cm from the source’s mid-point as described in our previous study 3,4. Low-energy photons emitted from the source encapsulation are suppressed in the air-kerma calculations by discarding all photons with energy less than 10 keV (i.e., PCUT set to 10 keV in EGSnrc). MC uncertainties are only statistical uncertainties (k=1).

Author Method Λ (cGy h-1 U-1) Abs. Uncertainty
Safigholi et al (with e- transport) 5 10x10x0.5 cm3 voxel at 100 cm 1.0985 0.0008
Safigholi et al (no e- transport) 5 10x10x0.5 cm3 voxel at 100 cm 1.0845 0.0002
Granero et al (no e- transport) 1 GEANT (extrap) 1.087 0.002
Selvam et al (with e- transport) 6 EGSnrc 1.097 0.002
Guerrero et al  (with e- transport) 2 PENELOPE 1.094 0.0003
Perez-Calatayud et al 7 Consensus value  1.092 0.005

Radial dose function - g(r):

The radial dose function, g(r), is calculated using both line and point source geometry functions and tabulated at 36 different radial distances ranging from 0.2 cm to 20 cm. For radii < 0.2 cm the values drop considerably. 5

 

Click image for higher res version

Note:  All the curves presented "with e- transport" do not include dose from the initial beta spectrum. For more details see our database paper  5 .


Anisotropy function - F(r,θ):

Anisotropy functions are calculated using the line source approximation and tabulated at 13 radii from 0.25 cm to 20 cm and 44 unique polar angles with a  resolution of 5° or better. The anisotropy factor, φan (r), was calculated by integrating the solid angle weighted dose rate over 0° ≤ ϑ ≤ 180°. For r < 0.5 cm, there can be significant differences with/without e- transport 5 .

 

Click images for higher res versions 
 F(1.00,θ) 
Anisotropy function
 F(5.00,θ) 
Anisotropy function
F(10.00,θ)
Anisotropy function

 

Along-Away Dose Data:

Along-away dose data are tabulated at 16 away distances from 0 cm to 20 cm and 29 along points from -20 cm to 20 cm. Doses are normalized to SK, the air-kerma strength.


Primary and Scatter Separated (PSS) Dose Data: D ii (r,θ):

Primary and Scatter Separated (PSS) dose data are tabulated at 12 radii from 0.3 cm to 20 cm and 47 unique polar angles with a  resolution of 5° or better. High resolution (Δr = 1 mm, ΔΘ = 1°) primary scatter dose data are also available in .csv files. For the purposes of these calculations, any photon escaping the source encapsulation is considered a primary. Only photons which scatter within the phantom are counted in the scatter tallies. Doses are normalized to the total photon energy escaping the encapsulation. The "ii" subscript labelled in the Dii(r, θ) represent the total scatter as Dto(r, θ), the primary photons as  Dpr(r, θ), the single scatter photon as  Dss(r, θ), and the multiple scatter photons as Dms (r, θ).

Click images for higher res versions
Dii (r,90°)*r 2
Primary and Scatter Separated (PSS) Dose Data
Dii (1.00,θ)
Primary and Scatter Separated (PSS) Dose Data
Dii (5.00,θ)
Primary and Scatter Separated (PSS) Dose Data
Dii (10.00,θ)
Primary and Scatter Separated (PSS) Dose Data

High resolution (1mm/1°) Tabulated Dii (r,θ) data in .csv format: Zipped archive 


Photon Energy Spectra

Photon energy spectra generated by the source model are calculated using the egs_brachy surface count scoring option to get the spectrum on the surface of the source. The plotted values are the counts per MeV in 1 keV bins, normalized to 1 count total in the spectrum. The MC calculations have a statistical uncertainty less than 0.001% on the mean energy.  The spectrum data are available in xmgrace format below.

Click image for higher res versions

Photon energy spectrum on the source surface: xmgrace


Tabulated data:

Tabulated data are available in .xlsx format: Excel


References:

1. D. Granero, J. Pérez-Calatayud, and F. Ballester, Technical note: Dosimetric study of a new Co-60 source used in brachytherapy, Med. Phys., 34 , 3485–3488, 2007
2. R. Guerrero et al, Dosimetric characterization of the 60Co BEBIG Co0.A86 high dose rate brachytherapy source using PENELOPE, Physica Medica, 30 , 960-967, 2014
3. R. E. P. Taylor et al , Benchmarking BrachyDose: voxel-based EGSnrc Monte Carlo calculations of TG--43 dosimetry parameters, Med. Phys., 34 , 445 - 457, 2007
4. D. W. O. Rogers, Inverse square corrections for FACs and WAFACs, Appl. Radiat. Isot.,153 ,108638, 2019
5. H. Safigholi, M. J. P. Chamberland, R. E. P. Taylor, M. P. Martinov, D. W. O. Rogers, and R. M. Thomson, Update of the CLRP TG-43 parameter database for high-energy brachytherapy sources,  to be published (Current calculation)
6. T. P. Selvam and S. Bhola, Technical note: EGSnrc-based dosimetric study of the BEBIG 60Co HDR brachytherapy sources, Med. Phys., 34, 1365–1370, 2010
7. Pérez-Calatayud et al , Dose Calculation for Photon-Emitting Brachytherapy Sources with Average Energy Higher than 50 keV: Full Report of the AAPM and ESTRO, 2012 by AAPM, ISBN: 978-1-936366-17-0