Radial dose function data for the Buchler G0814 source
Radius / cm | gL(r) L=0.13 cm | gp(r) | Unc. % |
0.20 | 1.034 | 1.001 | 0.05 |
0.25 | 1.023 | 1.002 | 0.07 |
0.30 | 1.016 | 1.002 | 0.08 |
0.40 | 1.010 | 1.003 | 0.11 |
0.50 | 1.008 | 1.003 | 0.13 |
0.60 | 1.002 | 0.999 | 0.16 |
0.70 | 1.007 | 1.006 | 0.19 |
0.75 | 1.005 | 1.004 | 0.20 |
0.80 | 1.004 | 1.003 | 0.22 |
0.90 | 1.001 | 1.000 | 0.24 |
1.00 | 1.000 | 1.000 | 0.27 |
1.25 | 1.007 | 1.007 | 0.05 |
1.50 | 1.008 | 1.008 | 0.06 |
1.75 | 1.010 | 1.011 | 0.07 |
2.00 | 1.011 | 1.012 | 0.08 |
2.50 | 1.011 | 1.012 | 0.10 |
3.00 | 1.013 | 1.015 | 0.12 |
3.50 | 1.015 | 1.016 | 0.14 |
4.00 | 1.013 | 1.015 | 0.16 |
4.50 | 1.013 | 1.014 | 0.18 |
5.00 | 1.007 | 1.009 | 0.20 |
6.00 | 1.005 | 1.006 | 0.11 |
7.00 | 0.995 | 0.996 | 0.13 |
8.00 | 0.984 | 0.985 | 0.15 |
9.00 | 0.968 | 0.970 | 0.17 |
10.00 | 0.951 | 0.953 | 0.18 |
11.00 | 0.933 | 0.934 | 0.10 |
12.00 | 0.911 | 0.912 | 0.11 |
13.00 | 0.888 | 0.889 | 0.12 |
14.00 | 0.864 | 0.865 | 0.13 |
15.00 | 0.838 | 0.839 | 0.14 |
16.00 | 0.810 | 0.811 | 0.15 |
17.00 | 0.783 | 0.784 | 0.16 |
18.00 | 0.753 | 0.755 | 0.17 |
19.00 | 0.727 | 0.728 | 0.18 |
20.00 | 0.695 | 0.696 | 0.19 |
Fitting coefficients for gL(r) = (a0r-2 + a1r-1 +a2 + a3r +a4r2 + a5r3) e-a6r | ||||||||
Fit range | Coefficients | |||||||
rmin (cm) | rmax (cm) | a0 / cm2 | a1 / cm | a2 | a3 / cm-1 | a4 / cm-2 | a5 / cm-3 | a6 / cm-1 |
0.20 | 20.00 | -1.3905E-03 | 2.1241E-02 | 9.5524E-01 | 1.6567E-01 | 6.5575E-03 | 5.0753E-04 | 1.3787E-01 |
Fitting coefficients for gL(r) = a0 + a1r +a2r2 + a3r3 +a4r4 + a5r5 | ||||||||
Fit range | Coefficients | |||||||
rmin (cm) | rmax (cm) | a0 | a1 / cm-1 | a2 / cm-2 | a3 / cm-3 | a4 / cm-4 | a5 / cm-5 | |
0.20 | 20.00 | 1.0221E+00 | -3.2081E-02 | 1.1394E-02 | -1.5279E-03 | 7.8589E-05 | -1.4378E-06 |