Radial dose function data for the PharmaSeed BT-125-2 source
| Radius / cm | gL(r) L=0.325 cm | gp(r) | Unc. % |
| 0.05 | 1.117 | 0.441 | 0.03 |
| 0.06 | 1.071 | 0.486 | 0.02 |
| 0.07 | 1.057 | 0.535 | 0.02 |
| 0.08 | 1.055 | 0.583 | 0.02 |
| 0.09 | 1.057 | 0.629 | 0.02 |
| 0.10 | 1.060 | 0.671 | 0.02 |
| 0.15 | 1.080 | 0.830 | 0.03 |
| 0.20 | 1.088 | 0.922 | 0.04 |
| 0.25 | 1.091 | 0.976 | 0.05 |
| 0.30 | 1.088 | 1.006 | 0.05 |
| 0.40 | 1.081 | 1.035 | 0.07 |
| 0.50 | 1.072 | 1.045 | 0.09 |
| 0.60 | 1.059 | 1.043 | 0.11 |
| 0.70 | 1.046 | 1.037 | 0.13 |
| 0.75 | 1.038 | 1.031 | 0.13 |
| 0.80 | 1.031 | 1.026 | 0.14 |
| 0.90 | 1.015 | 1.012 | 0.16 |
| 1.00 | 0.997 | 0.997 | 0.00 |
| 1.50 | 0.910 | 0.915 | 0.11 |
| 2.00 | 0.814 | 0.819 | 0.16 |
| 2.50 | 0.721 | 0.726 | 0.21 |
| 3.00 | 0.636 | 0.641 | 0.27 |
| 3.50 | 0.556 | 0.561 | 0.34 |
| 4.00 | 0.486 | 0.490 | 0.41 |
| 4.50 | 0.420 | 0.423 | 0.50 |
| 5.00 | 0.363 | 0.366 | 0.59 |
| 5.50 | 0.314 | 0.316 | 0.35 |
| 6.00 | 0.272 | 0.274 | 0.41 |
| 6.50 | 0.233 | 0.235 | 0.48 |
| 7.00 | 0.200 | 0.202 | 0.55 |
| 7.50 | 0.173 | 0.174 | 0.63 |
| 8.00 | 0.146 | 0.148 | 0.73 |
| 8.50 | 0.127 | 0.128 | 0.83 |
| 9.00 | 0.108 | 0.109 | 0.96 |
| 9.50 | 0.0918 | 0.0926 | 1.09 |
| 10.00 | 0.0790 | 0.0796 | 1.24 |
| Fitting coefficients for gL(r) = (a0r-2 + a1r-1 +a2 + a3r +a4r2 + a5r3) e-a6r | ||||||||
| Fit range | Coefficients | |||||||
| rmin (cm) | rmax (cm) | a0 / cm2 | a1 / cm | a2 | a3 / cm-1 | a4 / cm-2 | a5 / cm-3 | a6 / cm-1 |
| 0.05 | 10.00 | 4.8089E-04 | -1.1519E-02 | 1.1389E+00 | 5.0233E-01 | 2.7405E-03 | 4.4628E-03 | 4.9310E-01 |