Radial dose function data for the Advantage IA1-125A source
Radius / cm | gL(r) L=0.300 cm | gp(r) | Unc. % |
0.05 | 1.143 | 0.480 | 0.02 |
0.06 | 1.096 | 0.526 | 0.02 |
0.07 | 1.080 | 0.576 | 0.02 |
0.08 | 1.076 | 0.625 | 0.02 |
0.09 | 1.077 | 0.671 | 0.02 |
0.10 | 1.080 | 0.713 | 0.02 |
0.15 | 1.094 | 0.865 | 0.03 |
0.20 | 1.099 | 0.950 | 0.04 |
0.25 | 1.099 | 0.997 | 0.05 |
0.30 | 1.096 | 1.024 | 0.05 |
0.40 | 1.087 | 1.048 | 0.07 |
0.50 | 1.076 | 1.053 | 0.09 |
0.60 | 1.064 | 1.051 | 0.11 |
0.70 | 1.052 | 1.044 | 0.13 |
0.75 | 1.045 | 1.039 | 0.14 |
0.80 | 1.034 | 1.030 | 0.15 |
0.90 | 1.017 | 1.015 | 0.17 |
1.00 | 1.003 | 1.003 | 0.00 |
1.50 | 0.913 | 0.917 | 0.11 |
2.00 | 0.819 | 0.823 | 0.16 |
2.50 | 0.725 | 0.729 | 0.21 |
3.00 | 0.636 | 0.640 | 0.27 |
3.50 | 0.553 | 0.557 | 0.33 |
4.00 | 0.484 | 0.488 | 0.41 |
4.50 | 0.422 | 0.425 | 0.49 |
5.00 | 0.367 | 0.369 | 0.58 |
5.50 | 0.313 | 0.315 | 0.35 |
6.00 | 0.271 | 0.273 | 0.40 |
6.50 | 0.235 | 0.237 | 0.47 |
7.00 | 0.200 | 0.202 | 0.54 |
7.50 | 0.171 | 0.172 | 0.63 |
8.00 | 0.148 | 0.149 | 0.72 |
8.50 | 0.125 | 0.126 | 0.83 |
9.00 | 0.107 | 0.108 | 0.95 |
9.50 | 0.0919 | 0.0926 | 1.08 |
10.00 | 0.0802 | 0.0808 | 1.21 |
Fitting coefficients for gL(r) = (a0r-2 + a1r-1 +a2 + a3r +a4r2 + a5r3) e-a6r | ||||||||
Fit range | Coefficients | |||||||
rmin (cm) | rmax (cm) | a0 / cm2 | a1 / cm | a2 | a3 / cm-1 | a4 / cm-2 | a5 / cm-3 | a6 / cm-1 |
0.05 | 10.00 | 6.7921E-04 | -1.5741E-02 | 1.1828E+00 | 2.5623E-01 | -2.5011E-02 | 1.0344E-03 | 3.3650E-01 |